Lactate SCOUT - Tech Specs
Device type: Lactate SCOUT
Manufacturer: SensLab GmbH, Leipzig (Germany)
Measurement: enzymatic-amperometric detection of lactate in capillary blood
Measurement range: 0.5 - 25.0 mmol/l, automatic compensation of Hct influence
Sample volume: 0.5 µl
Measurement duration: 15 seconds
Coefficient of Var.: 3% - 8% (dependant on concentration)
Temperature: min. +5 °C, max. +45 °C, automatic control and compensation
Humidity: up to 85% relative humidity
Memory: 250 readings incl. additional data (time, storage place etc.)
Size: 91 x 55 x 24 mm
Weight: 3 oz with batteries
Display: Big LC-display with symbols
Power supply: 2 x 1.5V AAA/LR03-batteries (> 1.000 measurements)
Interface: Serial port with 2.5" socket, use original Senslab interface cable or Bluetooth adapter only!
Accesories: Lactate Scout-test strips, Lactate Scout-control solution

  1. Just pick a test strip out of the vial & insert into the Lactate Scout
  2. Clean by water, sanitize & dry the earlobe or fingertip carefully
  3. Prick the area for blood withdrawal (only a tiny droplet required)
  4. Hold the Lactate Scout with test strip onto the blood droplet
  5. Only 0.5 µl of blood will be absorbed, the measurement occurs automatically
  6. After 15 sec measuring duration, the result is displayed and stored
  7. Manifold options for gradual tests & Software analysis on the PC
 Sampling notes