| Device type: |
|
Lactate SCOUT |
|
| Manufacturer: |
|
SensLab GmbH, Leipzig (Germany) |
|
| Measurement: |
|
enzymatic-amperometric detection of lactate in capillary blood |
|
| Measurement range: |
|
0.5 - 25.0 mmol/l, automatic compensation of Hct influence |
|
| Sample volume: |
|
0.5 µl |
|
| Measurement duration: |
|
15 seconds |
|
| Coefficient of Var.: |
|
3% - 8% (dependant on concentration) |
|
| Temperature: |
|
min. +5 °C, max. +45 °C, automatic control and compensation |
|
| Humidity: |
|
up to 85% relative humidity |
|
|
| Memory: |
|
250 readings incl. additional data (time, storage place etc.) |
|
| Size: |
|
91 x 55 x 24 mm |
|
| Weight: |
|
3 oz with batteries |
|
| Display: |
|
Big LC-display with symbols |
|
| Power supply: |
|
2 x 1.5V AAA/LR03-batteries (> 1.000 measurements) |
|
| Interface: |
|
Serial port with 2.5" socket, use original Senslab interface cable or Bluetooth adapter only! |
|
| Accesories: |
|
Lactate Scout-test strips, Lactate Scout-control solution |
- Just pick a test strip out of the vial & insert into the Lactate Scout
- Clean by water, sanitize & dry the earlobe or fingertip carefully
- Prick the area for blood withdrawal (only a tiny droplet required)
- Hold the Lactate Scout with test strip onto the blood droplet
- Only 0.5 µl of blood will be absorbed, the measurement occurs automatically
- After 15 sec measuring duration, the result is displayed and stored
- Manifold options for gradual tests & Software analysis on the PC
Sampling notes |
|
|
|